Film ID: 73752
Presenting Author: Kei Ichikawa, MD
Producers: Kazuo Ichikawa, PhD, MD; Yukihito Kato; Yoshiki Tanaka, PhD; Seiji Tokiwa
Category: Instruments & Devices/IOLs
This study is find out the cause, we used intraoperative OCT to confirm the condition of the toric IOL during surgery. As a result, it was possible to observe that BSS remained between the toric IOL and the posterior capsule. In these cases, we succeeded to eliminate the BSS under the toric IOL by lowering the intraocular pressure.
Presenting Author: Kei Ichikawa, MD
Producers: Kazuo Ichikawa, PhD, MD; Yukihito Kato; Yoshiki Tanaka, PhD; Seiji Tokiwa
Category: Instruments & Devices/IOLs
This study is find out the cause, we used intraoperative OCT to confirm the condition of the toric IOL during surgery. As a result, it was possible to observe that BSS remained between the toric IOL and the posterior capsule. In these cases, we succeeded to eliminate the BSS under the toric IOL by lowering the intraocular pressure.